Skip to content
LexBuild

7 CFR § 801.6 - Tolerances for moisture meters.

---
identifier: "/us/cfr/t7/s801.6"
source: "ecfr"
legal_status: "authoritative_unofficial"
title: "7 CFR § 801.6 - Tolerances for moisture meters."
title_number: 7
title_name: "Agriculture"
section_number: "801.6"
section_name: "Tolerances for moisture meters."
chapter_name: "AGRICULTURAL MARKETING SERVICE (FEDERAL GRAIN INSPECTION SERVICE, FAIR TRADE PRACTICES PROGRAM), DEPARTMENT OF AGRICULTURE"
subchapter_number: "A"
subchapter_name: "FEDERAL GRAIN INSPECTION"
part_number: "801"
part_name: "OFFICIAL PERFORMANCE REQUIREMENTS FOR GRAIN INSPECTION EQUIPMENT"
positive_law: false
currency: "2026-03-24"
last_updated: "2026-03-24"
format_version: "1.1.0"
generator: "[email protected]"
authority: "7 U.S.C. 71-87k"
regulatory_source: "51 FR 7050, Feb. 28, 1986, unless otherwise noted."
cfr_part: "801"
---

# 801.6 Tolerances for moisture meters.

(a) The maintenance tolerances for Motomco 919 moisture meters used in performing official inspection services shall be:

(1) Headquarters standard meters:

|  |  |  |
| --- | --- | --- |
| Low | ±0.05 percent moisture, mean deviation from National standard moisture meter using Hard Red Winter wheat |  |
| Mid | ±0.05 percent moisture, mean deviation from National standard moisture meter using Hard Red Winter wheat |  |
| High | ±0.05 percent moisture, mean deviation from National standard moisture meter using Hard Red Winter wheat |  |

(2) All other than Headquarters standard meters:

|  |  |  |
| --- | --- | --- |
| Low | ±0.15 percent moisture, mean deviation from standard moisture meter using Hard Red Winter wheat | ±0.20 percent moisture, mean deviation from standard moisture meter using Hard Red Winter wheat |
| Mid | ±0.10 percent moisture, mean deviation from standard moisture meter using Hard Red Winter wheat | ±0.15 percent moisture, mean deviation from standard moisture meter using Hard Red Winter wheat |
| High | ±0.15 percent moisture, mean deviation from standard moisture meter using Hard Red Winter wheat | ±0.20 percent moisture, mean deviation from standard moisture meter using Hard Red Winter wheat |

(b) The maintenance tolerances for GAC 2100 moisture meters used in performing official inspection services shall be:

(1) Headquarters standard meters. By direct comparison using mid-range Hard Red Winter wheat, ±0.05% mean deviation for the average of the Headquarters standard moisture meters.

(2) All other than Headquarters standard meters. By sample exchange using mid-range Hard Red Winter wheat, ±0.15% mean deviation from the standard meter.

[63 FR 34554, June 25, 1998]