10 CFR § 35.2647 - Records of additional technical requirements for mobile remote afterloader units.
---identifier: "/us/cfr/t10/s35.2647"source: "ecfr"legal_status: "authoritative_unofficial"title: "10 CFR § 35.2647 - Records of additional technical requirements for mobile remote afterloader units."title_number: 10title_name: "Energy"section_number: "35.2647"section_name: "Records of additional technical requirements for mobile remote afterloader units."chapter_name: "NUCLEAR REGULATORY COMMISSION"part_number: "35"part_name: "MEDICAL USE OF BYPRODUCT MATERIAL"positive_law: falsecurrency: "2026-04-05"last_updated: "2026-04-05"format_version: "1.1.0"generator: "[email protected]"authority: "Atomic Energy Act of 1954, secs. 81, 161, 181, 182, 183, 223, 234, 274 (42 U.S.C. 2111, 2201, 2231, 2232, 2233, 2273, 2282, 2021); Energy Reorganization Act of 1974, secs. 201, 206 (42 U.S.C. 5841, 5846); 44 U.S.C. 3504 note."regulatory_source: "67 FR 20370, Apr. 24, 2002, unless otherwise noted."cfr_part: "35"---
Identifier
/us/cfr/t10/s35.2647
Currency
2026-04-05
Positive Law
No
Updated
2026-04-05
Chapter
Nuclear Regulatory Commission
Authority
Atomic Energy Act of 1954, secs. 81, 161, 181, 182, 183, 223, 234, 274 (42 U.S.C. 2111, 2201, 2231, 2232, 2233, 2273, 2282, 2021); Energy Reorganization Act of 1974, secs. 201, 206 (42 U.S.C. 5841, 58... Atomic Energy Act of 1954, secs. 81, 161, 181, 182, 183, 223, 234, 274 (42 U.S.C. 2111, 2201, 2231, 2232, 2233, 2273, 2282, 2021); Energy Reorganization Act of 1974, secs. 201, 206 (42 U.S.C. 5841, 5846); 44 U.S.C. 3504 note.
# 35.2647 Records of additional technical requirements for mobile remote afterloader units.(a) A licensee shall retain a record of each check for mobile remote afterloader units required by § 35.647 for 3 years.(b) The record must include—(1) The date of the check;(2) The manufacturer's name, model number, and serial number of the remote afterloader unit;(3) Notations accounting for all sources before the licensee departs from a facility;(4) Notations indicating the operability of each entrance door electrical interlock, radiation monitors, source exposure indicator lights, viewing and intercom system, applicators, source transfer tubes, and transfer tube applicator interfaces, and source positioning accuracy; and(5) The signature of the individual who performed the check.