21 CFR § 884.4150 - Bipolar endoscopic coagulator-cutter and accessories.

---
identifier: "/us/cfr/t21/s884.4150"
source: "ecfr"
legal_status: "authoritative_unofficial"
title: "21 CFR § 884.4150 - Bipolar endoscopic coagulator-cutter and accessories."
title_number: 21
title_name: "Food and Drugs"
section_number: "884.4150"
section_name: "Bipolar endoscopic coagulator-cutter and accessories."
chapter_name: "FOOD AND DRUG ADMINISTRATION, DEPARTMENT OF HEALTH AND HUMAN SERVICES"
subchapter_number: "H"
subchapter_name: "MEDICAL DEVICES"
part_number: "884"
part_name: "OBSTETRICAL AND GYNECOLOGICAL DEVICES"
positive_law: false
currency: "2026-04-05"
last_updated: "2026-04-05"
format_version: "1.1.0"
generator: "[email protected]"
authority: "21 U.S.C. 351, 360, 360c, 360e, 360j, 360, 371."
regulatory_source: "45 FR 12684, Feb. 26, 1980, unless otherwise noted."
cfr_part: "884"
---

# 884.4150 Bipolar endoscopic coagulator-cutter and accessories.

(a) *Identification.* A bipolar endoscopic coagulator-cutter is a device used to perform female sterilization and other operative procedures under endoscopic observation. It destroys tissue with high temperatures by directing a high frequency electrical current through tissue between two electrical contacts of a probe. This generic type of device may include the following accessories: an electrical generator, probes, and electrical cables.

(b) *Classification.* Class II. The special controls for this device are:

(1) FDA's:

(i) “Use of International Standard ISO 10993 ‘Biological Evaluation of Medical Devices—Part I: Evaluation and Testing,’ ”

(ii) “510(k) Sterility Review Guidance 2/12/90 (K-90),” and

(iii) “Guidance (‘Guidelines’) for Evaluation of Laproscopic Bipolar and Thermal Coagulators (and Accessories),”

(2) International Electrotechnical Commission's IEC 60601-1-AM2 (1995-03), Amendment 2, “Medical Electrical Equipment—Part 1: General Requirements for Safety,”

(3) American National Standards Institute/American Association for Medical Instrumentation's HF-18, 1993, “Electrosurgical Devices,”

(4) Labeling:

(i) Indication: For female tubal sterilization, and

(ii) Instructions for use:

(A) Destroy at least 2 centimeters of the fallopian tubes,

(B) Use a cut or undampened sinusoidal waveform,

(C) Use a minimum power of 25 watts, and

(D) For devices with ammeters: continue electrode activation for 5 seconds after the visual endpoint (tissue blanching) is reached or current flow ceases indicating adequate tissue destruction.