# 1065.274 Zirconium dioxide (ZrO) NO analyzer.

(a) *Application.* You may use a zirconia oxide (ZrO<sub>2</sub>) analyzer to measure NO<sub>X</sub> in raw exhaust for field-testing engines.

(b) *Component requirements.* We recommend that you use a ZrO<sub>2</sub> analyzer that meets the specifications in Table 1 of § 1065.205. Note that your ZrO<sub>2</sub>-based system must meet the linearity verification in § 1065.307.

(c) *Species measured.* The ZrO<sub>2</sub>-based system must be able to measure and report NO and NO<sub>2</sub> together as NO<sub>X</sub>. If the ZrO<sub>2</sub>-based system cannot measure all of the NO<sub>2</sub>, you may develop and apply correction factors based on good engineering judgment to account for this deficiency.

(d) *Interference.* You must account for NH<sub>3</sub> interference with the NO<sub>X</sub> measurement.