# 1065.295 PM inertial balance for field-testing analysis.
(a) *Application.* You may use an inertial balance to quantify net PM on a sample medium for field testing.
(b) *Component requirements.* We recommend that you use a balance that meets the specifications in Table 1 of § 1065.205. Note that your balance-based system must meet the linearity verification in § 1065.307. If the balance uses an internal calibration process for routine spanning and linearity verifications, the process must be NIST-traceable.
(c) *Loss correction.* You may use PM loss corrections to account for PM loss in the inertial balance, including the sample handling system.
(d) *Deposition.* You may use electrostatic deposition to collect PM as long as its collection efficiency is at least 95%.
[73 FR 59259, Oct. 8, 2008, as amended at 75 FR 68462, Nov. 8, 2010; 76 FR 57443, Sept. 15, 2011; 79 FR 23762, Apr. 28, 2014]